ProPulse Mushroom Valve & Washer Pack - 5 / Pack
The ProPulse Mushroom Valve is specifically designed to prevent water flowing out of the reservoir whilst filling.
Quantity - 5 / Pack
Ordering Information
- MHG-KIT5011 - Mushroom Valve & RED Washer Pack (G5, PP17, 2017) - 5 / Pack
- MHG-KIT1006 - Mushroom Valve & BLACK Washer Pack (II, III & NG) Older Units) - 5 / Pack
Should a replacement mushroom valve or washer be required, please see replacement steps as follows;
- Remove the reservoir from the ProPulse®
- Remove the old Mushroom Valve from the reservoir and discard
- Insert a new Mushroom Valve into the reservoir
- Remove the rubber Washer from the underside of the tank plug hole and discard
- Insert a new rubber Washer into the lip on the underside of the tank plug hole
- Return the reservoir to the machine
When the mushroom valve is fitted correctly, it should lift slightly with the tension of your finger, when pressed. Be careful when changing the valves and washers, as any damage to either will cause the tank to leak. Be sure to check the valves and washers prior to each use, as they do wear through normal use and may need changing – this will help prevent possible damage to the water inlet valve
A full range of essential ear care products that complement the ProPulse® Ear Irrigator are available through Stark Medical. These include;
- ProPulse® Single Use QrX™ Tips
- ProPulse® Cleaning Tablets
- ProScoop™ Ear Wax Removal Scoop
- ProTect™ Shoulder Cape
- ProPulse® HeadLamp
- ProPulse Mushroom Valves
Resources
- ProPulse 2017 User Manual
- Cleaning Instructions
- Warranty Guidelines
- Mushroom Valve User Guide
- Cleaning Tablet MSDS
Digital Inservice
Stark Medical has produced 3 videos to help our customers setup, use and maintain their ProPulse Ear Irrigator. Please see the series as follows or to watch on YouTube please - click here.
Part 1 - Set Up Instructions (2 min)
Part 2 - Instructions for Use (2.5 min)
Part 3 - Device Maintenance (2.5 min)